C8000 Battery Testing System User Manual
271
Field
Description
Compatible
Chemistry
Limits
Units
Additional
requirements
Typical Range(s)
SM-BUS
Imported
Registers
Min.
Max.
Notes:
terminated
when the
voltage drop
reaches the
pre-set rate).
Negative
slope delay
time
Duration before
a negative slope
is detected.
NiMH &
NiCD
0
20
C-Minute
----
----
No
Zero value is
valid and
indicates
Parameter is
"OFF"
Standby
voltage
----
SLA, Li,
LiPh
>
End of discharge
<=
Minimal
Voltage
V
----
(SLA)
2.15~2.45V/cell
No
----
(Li)
3.90~4.35V/cell
(LiPh)
3.20~3.65V/cell
Maximum
charge
voltage
The charging
voltage for
constant voltage
batteries
SLA, Li,
LiPh
>
Minimal Voltage
< Maximum
Voltage
V
Maximum charge
voltage of 45V
(SLA)
2.15~2.60V/cell
0x15
----
(Li)
3.90~4.35V/cell
(LiPh)
3.40~3.85V/cell
Charge
termination
point
Termination
Point for
Constant
Voltage
Batteries
SLA, Li,
LiPh
Rating<=100ah
0.055
<Charge
Current by
25mA
A
----
0.01~0.10C
No
----
Rating>100ah
0.200
Code Options