![Cadex C8000 User Manual Download Page 98](http://html1.mh-extra.com/html/cadex/c8000/c8000_user-manual_3250378098.webp)
C8000 Battery Testing System User Manual
86
3.1.9.1
Ports
The following ports on the PC firewall must be “allowed”.
TCP port 51935 must remain open.
The required ports are open by default in Windows 2000.
3.1.9.2
Other Firewall Software or Hardware
For other firewall software or hardware,
3.1.10
Selecting a TCP/IP Address for the C8000
3.1.10.1
Before Entering the Settings
Ask the Network Administrator for:
A static IP Address for each C8000
The network mask
If the C8000 will be used stand-alone without BatteryLab, then ignore all
references to TCP/IP settings for BatteryLab. The TCP/IP Settings on
the C8000 are disabled by default.
An IP address is a set of four numbers that identifies a single
computer or device uniquely on a TCP/IP network. Each of the
four numbers can range from 0 to 255. If a Network administrator
is unavailable, follow the instructions below on a computer to
determine a unique IP address for each C8000, other proceed to
step 2:
Note:
If the network system uses dynamic IP address assignment,
ensure that any IP addresses assigned, to a C8000, are exempted from the
Dynamic IP Address range of the server/router.
1.
Determine a unique IP address for each C8000 to be
connected to the local network.
Ping the network for each IP address (from the PC) to
determine that the IP address is not already in use on the
local network. If an IP address is “in use”, determine
another unique IP address. To ping an IP address, see
3.1.10.2 Ping the Network to Check for Available IP address
2. To enter a unique IP address for each C8000, follow the steps
2.11.5 Specify the TCP/IP Address on the C8000
Summary of Contents for C8000