![Cadex C8000 User Manual Download Page 43](http://html1.mh-extra.com/html/cadex/c8000/c8000_user-manual_3250378043.webp)
C8000 Battery Testing System User Manual
31
2.6.2
Clean the C8000
Caution
Before cleaning, turn off the power and unplug the AC
power cord.
Do not allow any liquids to enter the case under any
circumstances.
Clean the C8000 only with a damp cloth moistened
with mild soap and water. Allow to thoroughly dry.
For more information, see
2.6.3
Replace the Fuse
If the C8000 does not power up, make sure that the power cord is
properly inserted and that power is available.
If the C8000 still does not power up, the main (AC input) fuse may
need to be replaced.
Replace fuses only with fuses of the same type and rating:
Fuse 5x20mm T8A 250V, Cadex part number 52-546-0800.
Caution
Before replacing the fuse, unplug the power cord and
allow the C8000 to cool for at least one minute.
The C8000 contains
capacitors
that could retain a
charge after the AC power is disconnected.
For more information, see
To replace the main fuse:
1.
Remove all batteries and adapters, turn off the C8000, and
unplug the power cord from the AC Input.
Wait at least one minute before continuing.
2.
Using a flathead screwdriver, position the screwdriver at
the top of the fuse cover (beneath the AC Input) and pry
open the fuse cover.
3.
Carefully pull out the fuse.
Summary of Contents for C8000