STATION
saves any changes made to the program settings or the target capacity to
the battery station. These changes override any settings in the active C-code.
5.
Press
ENTER
.
6.
Press
or
to scroll to
YES
to accept changes and press
ENTER
.
7.
Press
ESC
twice to return to the global display.
Note
To use all of the settings from the active C-code, this option
must
be set to
C-
CODE
. Setting it to C-
CODE
will erase all settings saved to the stations.
Resuming Service After Power Failure
If there is a power failure, the back-up memory in the analyzer keeps service data
for one hour. If the power failure lasts less than an hour, battery service resumes
from where it left off. If the power failure lasts longer than one hour, the battery
resumes from the beginning. This time can be modified:
To change the Power Failure Resume Hours
1.
From the global display, press
MENU
.
2.
Press
or
to scroll to
SYSTEM SETUP
, press
ENTER
.
3.
Press
or
to scroll to
POWER FAILURE RESUME HOURS
, press
ENTER
.
4.
Press
or
to scroll to the required setting:
5.
Press
ENTER
. Press
ENTER
again to save the settings.
6.
Press
ESC
twice to return to the global display.
81