To set HyperTerminal to read Cadex analyzer data
1.
Click
Start
on the Windows Task Bar, point to
Programs
, then point to
Accessories
, and then click
HyperTerminal
.
2.
Double-click the
HyperTrm
or
HyperTrm.exe
icon.
3.
Type a name for the data log connection, such as Cadex analyzer
connection, and then click
OK
.
4.
In the dialog box that appears, enter these settings:
Connect using
. From the list, select
Direct to Comx
, where x is the number of the
serial port to which the analyzer is connected.
•
Bits per second: 9600
•
Data Bits: 8
•
Parity: None
•
Stop Bits: 1
•
Flow Control: None
5.
Click
OK
.
The HyperTerminal screen appears. The bottom left-hand corner of the screen
should display the message “Connected h:mm:ss”.
6.
Check that HyperTerminal is communicating with the Cadex analyzer
by inserting an adapter in any station on the analyzer. A text string
similar to this one should appear on-screen:
0,2," ","01/25/2001","095126",201,"0\80"
If nothing appears on-screen, check the HyperTerminal properties:
Click
File
, and then click
Properties
.
Verify that the serial communication port number is correct.
Click
Configure
, and verify that the serial port settings match the Cadex analyzer
serial port settings.
97