Chapter 12
System Calibration
There are two kinds of calibration that should be performed on the Cadex analyzer
on a regular basis: voltage calibration and adapter calibration.
Voltage Calibration
Voltage calibration should be performed yearly and a calibration kit is required.
The C7000-C Series of Battery Analyzers uses Calibration Kit P/N 92-770-
0212 (See Appendix D -
To calibrate the voltage
1.
Remove all Battery Adapters.
2.
From the global display, press
MENU
.
3.
Press
or
to scroll to
SYSTEM SETUP
and then press
ENTER
.
4.
Press
or
to scroll to
CALIBRATE VOLTAGE
and then press
ENTER
.
5.
Firmly insert the four voltage calibration adapters, one into each
station. (If calibrating a C7200, only two voltage calibration adapters
are necessary)
6.
Wait for the prompt, and then remove the calibration adapters.
7.
Press
MENU
to return to the global display.
8.
Press
FN
, and then press
7
. The “#” sign appears in the upper right
corner of the startup screen, indicating successful completion of
voltage calibration.
Cadex Electronics Inc. C7000-C Series Battery Analyzers
.
112