Detailed
Message
Code Reasons
Solutions
BATTERY TOO
HOT
13
Battery temperature is too
hot when inserted. The
battery automatically starts
processing when
temperature drops 5°C
below maximum value set in
the C-code Temperature
Sensing.
Allow the battery to cool or
use the analyzer in a cooler
room.
If the battery is non-OEM, the
thermistor may not match that
on the adapter.
BATTERY
UNDERCHARGE
D
175
Battery SoC (State-of-
charge) is too low for
QuickTest or OhmTest
The battery is charged for
five minutes, after which
QuickTest restarts. Charge
is repeated up to 5 times
until SoC is above 20%.
If OhmTest is running, the
program is aborted. Partially
charge the battery before
running OhmTest again.
BATTERY
UNDERCHARGE
D
177
Current program completed;
battery SoC (State-of-
charge) low.
Charge the battery.
CALIBRATION
FAULT
170
Current cannot pass
through shorted terminal
during calibration.
A minimum AWG 10 (2.5 mm)
wire is recommended to short
contacts/leads.
The wires may have
disconnected during the
calibration process.
Repeat calibration procedure.
CAP
.
IMPROVED
TO TARGET
195
Target capacity is achieved
or exceeded after an initial
warning
See code 115, Target
capacity not met.