Program and target capacity
. This field displays a one- or two-digit number
representing the current program and the target capacity from the C-code. A
back-slash character separates the two elements.
Program codes are:
0
= Auto
9 = Self-Discharge
1
= Charge
10 = LifeCycle
2
= Prime
11 = Discharge Only
3
= Custom1
12 = Extended Prime
4
= Custom2
13 = Boost
5
= Custom3
14 = QuickTest
6
= Custom4
15 = Q-Learn
7
= OhmTest
16 = Learn
8
= Runtime
17 = QuickSort
•
Battery data
. In certain cases, event strings also contain a battery data field.
For example, when an automatic OhmTest is run as part of a program, the
event string generated includes the OhmTest results in m
Ω
.
Sample text strings
Here are some sample data logging text strings followed by explanations.
0,2," ","01/24/2001","085120",201,"0\80"
a) A Battery Adapter was inserted (Code 201) in station 2 on January
24, 2001 at 8:51 a.m. The current C-code specifies the Auto
program with a target capacity of 80% (0\80 in the final field).
Battery ID is not yet available.
0,2," ","01/24/2001","085121",20,"0\80"
b) A battery was inserted in the adapter (Code 20).
Cadex Electronics Inc. C7000-C Series Battery Analyzers
.
94