Detailed
Message
Code Reasons
Solutions
C-
CODE NOT
USABLE
214
C-code is not suitable or is
out of range for the
analyzer.
The C-code may have been
created on an “Extended
Range” (ER) analyzer or an
analyzer with more recent
firmware version.
Verify required battery
parameters and edit or re-
enter the C-code.
Update the firmware on the
analyzer to the most recent
version (see
CELL MISMATCH
112 For NiCd or NiMH batteries
only. Cells are reaching full
charge at different times so
the negative slope is not
well defined. The program
proceeds to the next cycle.
Wait until processing is
complete because the Auto
and Prime programs will
attempt to correct this warning
through several more charge
and discharge cycles. If
corrected, code 192 (Cell
mismatch corrected) appears.
Reduce the negative slope in
extended C-code to 8mV/cell.
Battery is old or has been in
storage. Operating time may
be reduced. Either monitor
battery use or discard the
battery.
If the battery is new, use the
Prime program.
CELL MISMATCH
CORRECTED
192
Cell mismatch (code 112) is
corrected through several
charge/discharge cycles.
Use the battery as normal.
Cadex Electronics Inc. C7000-C Series Battery Analyzers
.
134