Detailed
Message
Code Reasons
Solutions
STATION OFF
LINE
0
Station is not reading the
adapter.
Remove the adapter and
restart the analyzer.
Make sure all analyzers have
the same firmware version.
Delete the C-code that was
selected for the battery.
Reset the system, (see
Resetting the System
,
page 110).
SYSTEM TEMP
.
HIGH
:
COOLING
207
Service on all stations is
temporarily suspended due
to high temperature inside
the analyzer.
Service will resume in a few
minutes, after the board has
cooled down. If it reoccurs
continually, move the analyzer
to a cooler room.
The firmware chip in the
analyzer may be loose.
Contact Cadex for instructions
to reinsert firmware.
TARGET
CAPACITY NOT
MET
115
Battery capacity is below
the target capacity. The
analyzer attempts to
improve the capacity by
reconditioning the battery.
Wait until service is complete.
The Auto and Prime programs
will try to correct this warning.
If corrected, Code 195
appears. If not corrected,
Code 116 appears. See Code
195 and 116 for further
information.
TARGET
CAPACITY NOT
MET
116
Final capacity of the battery
is below the target capacity.
Attempts to improve battery
capacity did not succeed in
bringing capacity above
target.
Confirm that the battery rating
matches the C-code mAh
setting.
Battery is old and has less
operating time than
manufacturer’s specification.