Detailed
Message
Code Reasons
Solutions
PASSWORD
ENTERED
203
System password has been
entered and activated.
Password is required for some
actions, depending on security
level selected, (see
Security
,
page 74).
PLATEAU
TIMEOUT
113
Battery was fully charged
before full-charge conditions
were met. For SLA, Li –Ion
and Li-Ph only.
The end-of-charge setting in
the C-code may have to be
increased. Battery may be old
and operating time less than
manufacturer’s specifications.
POWER ON
200 Power is detected inside the
analyzer.
PROCESS
COMPLETE
15,
35, 36
Current battery-service
program is complete.
PROCESS
RESUMING
25
(Code 25) Battery service is
continuing after a power
interruption.
This message appears after
the initial startup screen and
remains until the analyzer
determines the point at which
battery service was
interrupted and resumes the
current program from that
point.
PROCESS
SUSPENDED
18
Battery is removed during
service.
Re-insert battery within 5
seconds to resume service. If
message persists, see Code
188 for further information.
RAPID HEAT
RISE
152
On initial charge, the battery
temperature is rising too
rapidly (2°C in 1 minute) For
safety, charging is
terminated.
Use ExtPrime program if the
battery has been in storage
Reduce the charge rate in the
extended C-code.
The battery may be old or
have shorted cells. Discard
the battery.
141
Summary of Contents for C7200
Page 2: ......