Detailed
Message
Code Reasons
Solutions
CURRENT RISE
AT FULL
CHARGE
130
Ffor SLA and Li-ion and Li-
Ph only. Battery current
suddenly increased as the
end-of-charge condition was
achieved.
Raise the End of Charge
settings.
If the SLA battery is a Hawker
cell, use the Hawker C-code
settings (see page 60). Allow
the station to complete
service.
CUSTOM
PROGRAM HAS
FAILED
–
DISPLAY WILL
SHOW REASON
FOR FAILURE
16
(Code 16) The custom
program has produced an
error. The Next statement is
not processed. The program
has been terminated.
See the detailed display for
the exact nature of battery
failure.
If Code 16 is displayed, then
there is a bug in the custom
program - verify the custom
program statements.
CYCLE
RESUMED
32
Power resumed or the
battery is re-inserted into
the adapter.
DCHG CYCLE
COMPLETE
31
Discharge cycle completed. Program is going to the next
step.
DISCHARGE
CURRENT LOW
162
The analyzer cannot
discharge the battery. Very
low values of current (<
50mA) are displayed on the
analyzer before this code
appears
(See code 160, Bad fuses or
driver).
DISCHARGE
TIMEOUT
142
Battery capacity has
exceeded 250% of the rated
capacity.
Verify that the mAh rating of
the battery matches the C-
code mAh setting.
Raise the discharge rate in
the extended C-code.
DISCHARGE
WAIT
6
Station is on hold until
sufficient power is available.
Wait until other stations have
completed battery service.