Visual-ICE™ Cryoablation System
7-1
7 TROUBLESHOOTING
Galil Medical suggests the following options for troubleshooting the Visual-ICE™ System. If the
suggested approaches do not resolve the issue, contact Galil Medical Customer Service.
7.1 Software Recovery
In the event of software corruption or failure, the software may be restored to the previous software
version. Administrative and service users may update the software with an appropriate USB flash
drive.
1. Shutdown the Visual-ICE System.
2. Hold down the
Software Reset
button by inserting a straightened paper clip into the Software
Reset hole while simultaneously powering up the system. The system displays the
Software
Recovery Screen
.
Screen 7-1. Software Recovery Screen
3. Press the
Rollback
button to restore the software to the previous software version.
4.
OPTIONAL:
Press the
Load
button to update the software if the screen displayed a message
indicating that the software configuration is invalid on the login screen (Screen 7-2).