Visual-ICE™ Cryoablation System
4-3
The diagnostic tests check:
•
The correct version of firmware is running on the system
•
Critical system components including the solenoid valves, internal power supplies, cooling
fans, pressure transducers, and temperature measuring circuits.
If the system detects a failure of an individual channel, the channel is disabled and the needle
type display window, bordered in red, notes the channel is defective. A message indicating the
failure is displayed in the upper right-hand corner of the Navigation Tool Bar
.
Screen 4-1. Disabled Channel
A failure that prevents use of the system displays a message directing you to contact Galil
Medical Customer Service (reference Section 7.7).
If the Visual-ICE System software detects pressurized gas in the system and the gas supply is
not connected, a message appears requesting you to vent the gas from the system.
Screen 4-2. Vent Gas Message
Summary of Contents for Visual-ICE
Page 1: ...User Manual ...
Page 10: ...viii User Manual This page intentionally left blank ...
Page 22: ...1 10 User Manual This page intentionally left blank ...
Page 30: ...2 8 User Manual This page intentionally left blank ...
Page 76: ...4 30 User Manual This page intentionally left blank ...
Page 114: ...7 32 User Manual This page intentionally left blank ...
Page 122: ...9 2 User Manual This page intentionally left blank ...
Page 124: ...10 2 User Manual This page intentionally left blank ...
Page 125: ......