Visual-ICE™ Cryoablation System
3-5
Button
Description
Report
Displays a report of all procedure data that has been entered
and captured for the current procedure. The report can be
saved to the Galil Medical USB flash drive. Pressing the
Report
button during a procedure displays all procedure
information saved up to that point in time.
End Procedure
Ends the current procedure and returns to the
Startup Screen
.
Pressing this button generates a confirmation request, a
request to save the report, and an option to automatically vent
the system.
3.3.2 Context Sensitive Help
The
Context Sensitive Help
displays a summary of procedural steps to guide you through a
cryoablation procedure. It is intended to provide only a general overview; step-by-step instructions
3.3.3 User-Selected Self Help
The title bar of each section provides access to additional help information. Press the title bar to
access an explanation of the buttons and fields available in each section of the
Procedure Screen
.
3.3.4 Channel Controls
Channels 1 through 10 are labeled individually and contain independent controls for
Test, Freeze,
Freeze Intensity, Thaw,
and
Stop
. Each individual channel displays the needle type of the needles
connected adjacent to the channel controls (Screen 3-5). The Channel labeled
ALL
operates all
active channels simultaneously.
Table 3-3. Channel Controls
Button
Description
Channel
button – Identifies the active channel(s).
Advanced Channel Controls:
Pressing and holding the
Channel
button provides options to change the needle
type selected for that channel, link two adjacent channels
together for simultaneous operation, and program freeze –
thaw cycles.
Channel labeled
ALL
– Allows testing, freezing, and thawing
on ALL active channels simultaneously. Press the button for
the desired function (
Test, Freeze,
or
Thaw
) on this channel to
activate that function on all needles simultaneously.
Test
button – Begins the Needle Integrity and Functionality
Testing that is required prior to using any cryoablation needle.
No other controls are enabled until needle testing is complete.
Summary of Contents for Visual-ICE
Page 1: ...User Manual ...
Page 10: ...viii User Manual This page intentionally left blank ...
Page 22: ...1 10 User Manual This page intentionally left blank ...
Page 30: ...2 8 User Manual This page intentionally left blank ...
Page 76: ...4 30 User Manual This page intentionally left blank ...
Page 114: ...7 32 User Manual This page intentionally left blank ...
Page 122: ...9 2 User Manual This page intentionally left blank ...
Page 124: ...10 2 User Manual This page intentionally left blank ...
Page 125: ......