Visual-ICE™ Cryoablation System
7-29
Message
Reason for Appearance
80-31 System Error
•
Gas cylinder shut off valve is not open enough to
provide adequate flow. If necessary, open the valve
approximately another half turn.
•
Galil Medical Customer Service
o +1 877 639 2796 or
o +1 651 287 5000
o
Gas flow from gas cylinder was
insufficient. To improve gas flow, the
cylinder valve should be opened further.
80-32 System Error
•
Channel X is defective. Choose another channel.
At the conclusion of the procedure, contact Galil
Medical Customer Service.
•
Galil Medical Customer Service
o +1 877 639 2796 or
o +1 651 287 5000
o
A defective solenoid on Channel X was
detected; another channel should be
selected.
80-33 System Error
•
Defective MTS in channel [X]. Replace with a new
MTS.
•
Galil Medical Customer Service
o +1 877 639 2796 or
o +1 651 287 5000
o
The MTS temperature display was out
of the expected range when initially
connected to the needle connection
panel.