7-30
User Manual
Message
Reason for Appearance
80-34 System Error
•
Fan X failure. At the conclusion of the procedure,
contact Galil Medical Customer Service.
•
Galil Medical Customer Service
o +1 877 639 2796 or
o +1 651 287 5000
o
Fan X was identified as not operating.
80-35 System Error
•
The gas flow rate in channel XX exceeds the
system operation requirements and may impact
performance. Reduce the number of active needles.
•
Galil Medical Customer Service
o +1 877 639 2796 or
o +1 651 287 5000
o
The calculated flow rate for a given
channel exceeded the optimum system
requirement. The number of active
needles should be reduced.
80-36 System Error
•
The gas flow rate exceeds the system operation
requirements and may impact performance. Reduce
the number of active needles
•
Galil Medical Customer Service
o +1 877 639 2796 or
o +1 651 287 5000
o
The cumulative flow rate for all channels
exceeded the optimum system
requirement. The number of active
needles should be reduced.