5.4 - Service software
M2007/3en
TCT-2301 Service instructions
105
©BTG 2004
Reply time out
The instrument doesn’t
answer on command. (Waiting
time 256 ms).
SPC is probably not
connected to the instrument
or the instrument isn’t
switched on.
Also check the cables be-
tween SPC and the instru-
ment.
HART rev error
Wrong version of the HART
protocol. Should be version 5.
Update the software in SPC
or the instrument so that
they can understand one
another.
Not
TCT-2300
dev
Wrong type of instrument.
(
TCT-2300
is the expected
type of instrument, that is, it is
controlled from the application
started in SPC).
Select the correct program
in SPC or the correct instru-
ment to connect it to.
TCT-2300
rev err
Correct type of instrument (in
this case a TCT-2300 is ex-
pected), but the protocol ver-
sion is not compatible.
Update the software in SPC
or the instrument so that
they can understand one
another.
Dev change/repl
Another instrument is connect-
ed than that which the data
comes from. Identified by the
serial number.
Connect to the correct in-
strument and start by READ
DATA.
If the instrument is the cor-
rect one, continue to per-
form READ DATA.
Error message
Probable cause
Solution