3.2 - Installing the transmitter
M2007/3en
TCT-2301 Installation instructions
25
©BTG 2004
2. Apply a small amount of o-ring lubricant to the o-ring in the ball
valve housing.
3. Make sure the o-ring on the probe flange is mounted.
4. Insert the probe, using a twisting motion to get the probe tip to enter
the o-ring. Place the flexible insertion screw in the correct position
and make sure the hook with its locking device is in the correct po-
sition and secured.
5. Open the ball valve and insert the probe by cranking the handle.
6. Secure the probe in inserted position with the two security screws
!
Secure the two security screws properly
7. When removing the probe, reverse the above procedures.
Fig 16 Mounting the probe in
the valve assembly
Avoid getting O-ring
lubricant on the measuring
gap