User Manual
5 Step-by-Step Guides
94
5.13.3
Standard-based Zeta Quantification
This section describes how to perform standard-based quantification of electron transparent
samples using beam current and thickness of standard sample as reference values.
Step
Example/hints
1
Click the
Standards configurator
’s
calibration icon
to open the
SYSTEM FACTOR CALIBRATION
dialog.
2
Measure and enter the beam current
value as system factor.
Follow section 5.10 for electron transparent
samples. Enter the beam current in [pA] units.
3
Check or Load the appropriate Zeta
factors.
Follow section 5.11.3.
4
Select the
or
workspace and
a)
a spectrum
b) use the Spectrum chart
icon to
open a spectrum
Enter the beam current as spectrum
parameter: double click on the highlighted line
of the active spectrum in the spectrum list
(below the spectrum) to open the
SPECTRUM PROPERTIES
dialog and enter
the beam current in pA as system factor in
the
Parameter
tab.
c) drag and drop a spectrum from project
to Spectrum chart.
5
Click on the
button and
Load…
a quantification method.
Select a TEM quantification method.
Alternatively drag and drop a quantification
method from the project to spectrum.
6
Set the quantification method to
interactive
(
).
7
Click on the
button to
access the
QUANTIFICATION
dialog.
8
Set
Quantification model
to Zeta factor
method and perform quantification of
“unknown” sample.
See section 5.12.1 to optimize quantification
and to save or export quantification method
and results.
Calculated thickness is shown below the
result table. Density is calculated based on
element composition. If density is known, it
can be entered and sample thickness will
change correspondingly.
9
Click
to close the
QUANTIFICATION
dialog.
Summary of Contents for QUANTAX EDS
Page 1: ...QUANTAX EDS Energy dispersive X ray spectrometer for electron microscopy User Manual ...
Page 6: ......
Page 8: ...User Manual 2 Safety Information 8 ...
Page 12: ......
Page 14: ...User Manual 3 The QUANTAX EDS System 14 ...
Page 22: ......
Page 24: ......
Page 48: ......
Page 50: ...User Manual 5 Step by Step Guides 50 ...
Page 54: ...User Manual 5 Step by Step Guides 54 Fig 5 2 1 Spectrometer calibration ...
Page 56: ...User Manual 5 Step by Step Guides 56 Fig 5 3 1 Image capture ...
Page 58: ...User Manual 5 Step by Step Guides 58 Fig 5 4 1 Image extension ...
Page 60: ...User Manual 5 Step by Step Guides 60 Fig 5 5 1 Image drift correction ...
Page 62: ...User Manual 5 Step by Step Guides 62 Fig 5 6 1 Image processing ...
Page 67: ...5 8 Using the Spectrum Chart QUANTAX EDS 67 Fig 5 8 1 Features in the spectrum chart ...
Page 72: ...User Manual 5 Step by Step Guides 72 Fig 5 10 2 Beam current setting ...
Page 75: ...5 11 Standards Library QUANTAX EDS 75 Fig 5 11 2 Quantification dialog ...
Page 97: ...5 14 Object Analysis QUANTAX EDS 97 Fig 5 14 1 Object analysis ...
Page 108: ...User Manual 5 Step by Step Guides 108 Fig 5 19 1 Map processing ...
Page 111: ...5 20 Phase Analysis QUANTAX EDS 111 Fig 5 20 1 Performing phase analysis ...
Page 113: ...5 21 Creating a Report QUANTAX EDS 113 Fig 5 21 1 Creating a report ...
Page 116: ......