4.5 Configurator Bar
QUANTAX EDS
35
Fig. 4.5-12 Scan configuration dialog
The dwell time can only be set
in increments of numbers which are a
power of two. Direct input of arbitrary
numbers is not possible. Use the
arrows in the input field.
Very high image resolution will
slow down operation.
For
Imaging
,
Mapping
and
Line scan
different timing settings are
maintained. However, the electron
image captured in the left image
window of each workspace always
follows the settings for
Imaging
.
The scan speed can also be
set from the dialog that opens after
clicking the
icon next to the
image chart.
desired detector can be selected here. For each
channel only one image detector can be selected at a
time. If only Ch1 and Ch2 can be selected from the
dropdown menu, note that the image channel names
can be changed in the System workspace under
Microscope to e.g. SE, BSE, HAADF,… (refer to
section 4.7.6).
Imaging.
The
Dwell time
is the time the electron
beam stays on a pixel while capturing an image.
Select expanded dwell times to allow filtering and
averaging of the image signal and obtain less noisy
images. The
Line average
factor controls the
number of times a horizontal line is scanned in the
slow scan direction and averaged before proceeding
to the next line. Increasing this factor provides noise
reduction and expanded image scan time while
limiting the increased stress on the sample by
maintaining a high scan speed. The line average
factor will multiply the total image acquisition time.
The total
Image capture time
is calculated and
displayed.
Mapping.
The
Dwell time
is the time the electron
beam stays on a pixel while acquiring an EDS map.
The
Line average
factor for mapping controls the
number of times a horizontal line is scanned in the
slow scan direction and averaged before proceeding
to the next line. The capture time (
Mapping scan
time
) for one mapping frame is calculated by
multiplying the numbers of pixels and lines with the
dwell time and the line average factor. The real scan
time may be longer due to the time needed for data
processing by the computer.
Drift correction.
When clicking on the
Drift
correction
button, the
IMAGE DRIFT CORRECTION
SETTINGS
dialog pops up. It is described in detail in
section 5.5.
Power synchronization.
When the according
control box is checked, the scan process is
synchronized to the cycles of the AC mains voltage.
Power synchronization substantially reduces blurring
of vertical lines caused by electromagnetic
interference (at the expense of the scan speed). It is
applicable for capturing any image or map. However,
it is normally only useful in connection with very high
magnification of the microscope. A special AC/AC
adapter to be connected to the QUANTAX server is
provided.
Hardware inputs.
The IO scan card provides 8
Summary of Contents for QUANTAX EDS
Page 1: ...QUANTAX EDS Energy dispersive X ray spectrometer for electron microscopy User Manual ...
Page 6: ......
Page 8: ...User Manual 2 Safety Information 8 ...
Page 12: ......
Page 14: ...User Manual 3 The QUANTAX EDS System 14 ...
Page 22: ......
Page 24: ......
Page 48: ......
Page 50: ...User Manual 5 Step by Step Guides 50 ...
Page 54: ...User Manual 5 Step by Step Guides 54 Fig 5 2 1 Spectrometer calibration ...
Page 56: ...User Manual 5 Step by Step Guides 56 Fig 5 3 1 Image capture ...
Page 58: ...User Manual 5 Step by Step Guides 58 Fig 5 4 1 Image extension ...
Page 60: ...User Manual 5 Step by Step Guides 60 Fig 5 5 1 Image drift correction ...
Page 62: ...User Manual 5 Step by Step Guides 62 Fig 5 6 1 Image processing ...
Page 67: ...5 8 Using the Spectrum Chart QUANTAX EDS 67 Fig 5 8 1 Features in the spectrum chart ...
Page 72: ...User Manual 5 Step by Step Guides 72 Fig 5 10 2 Beam current setting ...
Page 75: ...5 11 Standards Library QUANTAX EDS 75 Fig 5 11 2 Quantification dialog ...
Page 97: ...5 14 Object Analysis QUANTAX EDS 97 Fig 5 14 1 Object analysis ...
Page 108: ...User Manual 5 Step by Step Guides 108 Fig 5 19 1 Map processing ...
Page 111: ...5 20 Phase Analysis QUANTAX EDS 111 Fig 5 20 1 Performing phase analysis ...
Page 113: ...5 21 Creating a Report QUANTAX EDS 113 Fig 5 21 1 Creating a report ...
Page 116: ......