User Manual
5 Step-by-Step Guides
86
5.12.3
Standard-based Quantification with a Temporary Reference
This guide describes how to perform standard-based quantification without calibration of system
factor, measuring reference and unknown samples with same microscope/beam conditions.
Step
Example/hints
1
Prepare an “unknown” sample,
reference sample(s) and a sample for
system factor calibration.
Select workspace.
“Unknown” sample, reference sample(s) and
sample for system factor calibration must be
available at the same time in the SEM
chamber.
2
Calibrate the system factor.
For details see section 5.10. Do not change
microscope parameters afterwards.
3
Select the
or
workspace.
4
Move the reference sample into
analytical position and
a
spectrum.
Follow 5.7 to acquire a spectrum. Use
automatic spectrum numbering or rename
spectrum after acquisition.
5
Click at the
button and
select
Load…
Load a quantification method.
6
Set
Method mode
to Interactive (
).
7
Click
.
The
QUANTIFICATION
dialog opens.
Set quantification parameters (Element ID,
EDS background, Deconvolution) as needed.
8
Click
.
EDIT STANDARD PROPERTIES
dialog opens
9
Enter reference concentrations and click
.
NOTE:
The reference values are
TEMPORARILY stored until the ESPRIT
software is closed. They will not be added
into the standards library. For advanced
standard-based quantification options see
section 5.12.4.
10
The
VALIDATION
dialog pops up. Close
it with
.
11
Click
to close the
QUANTIFICATION
dialog.
12
Repeat steps 4-11 for other reference
samples.
13
Move the “unknown” sample into the
analytical position.
Do not change any relevant microscope
settings (beam current, focus, high voltage...)
14
a spectrum.
If automatic quantification during or after
acquisition is enabled (see 5.7), the spectrum
is quantified automatically and the results
appear in the result list.
Summary of Contents for QUANTAX EDS
Page 1: ...QUANTAX EDS Energy dispersive X ray spectrometer for electron microscopy User Manual ...
Page 6: ......
Page 8: ...User Manual 2 Safety Information 8 ...
Page 12: ......
Page 14: ...User Manual 3 The QUANTAX EDS System 14 ...
Page 22: ......
Page 24: ......
Page 48: ......
Page 50: ...User Manual 5 Step by Step Guides 50 ...
Page 54: ...User Manual 5 Step by Step Guides 54 Fig 5 2 1 Spectrometer calibration ...
Page 56: ...User Manual 5 Step by Step Guides 56 Fig 5 3 1 Image capture ...
Page 58: ...User Manual 5 Step by Step Guides 58 Fig 5 4 1 Image extension ...
Page 60: ...User Manual 5 Step by Step Guides 60 Fig 5 5 1 Image drift correction ...
Page 62: ...User Manual 5 Step by Step Guides 62 Fig 5 6 1 Image processing ...
Page 67: ...5 8 Using the Spectrum Chart QUANTAX EDS 67 Fig 5 8 1 Features in the spectrum chart ...
Page 72: ...User Manual 5 Step by Step Guides 72 Fig 5 10 2 Beam current setting ...
Page 75: ...5 11 Standards Library QUANTAX EDS 75 Fig 5 11 2 Quantification dialog ...
Page 97: ...5 14 Object Analysis QUANTAX EDS 97 Fig 5 14 1 Object analysis ...
Page 108: ...User Manual 5 Step by Step Guides 108 Fig 5 19 1 Map processing ...
Page 111: ...5 20 Phase Analysis QUANTAX EDS 111 Fig 5 20 1 Performing phase analysis ...
Page 113: ...5 21 Creating a Report QUANTAX EDS 113 Fig 5 21 1 Creating a report ...
Page 116: ......