5.2 Energy-Channel Calibration
QUANTAX EDS
53
5.2
Energy-Channel Calibration
Correct energy-channel calibration is a prerequisite for reliable qualitative and quantitative analysis
results. Calibration should be checked on a regular basis or after re-start of the signal processing
unit (SVE).
Step
Examples/hints
1
Select the
workspace and
the
Spectrometer
tab.
Do not confuse energy-channel calibration
with system factor calibration (relevant for
standard-based quantification; for details see
section 5.12.3 and 5.12.4).
2
Set a calibration sample into the analysis
position.
Use a sample with any element in the mid
energy range (Mn – Zn) without peak overlap
and sufficient concentration. Preferably use
single element standards (e.g. pure copper) to
avoid peak misidentification, improve speed
and accuracy.
3
Set the beam current to produce an
intermediate count rate.
Use microscope controls. Set appropriate
magnification on a homogeneous sample
area.
4
Select element and spectral line for
calibration.
Use K
α
lines whenever possible.
5
Click
to check the current
energy channel calibration and deviation.
The test corresponds to the current
Pulse
throughput
and
Maximum energy
displayed
in the
EDS configurator
. The acceptable
deviation value is b 5 eV and - 5 eV.
If not, proceed with step 6.
6
Select
Pulse throughput
and
Energy
range
under
Settings to calibrate
.
7
Choose accuracy and click
.
Recommended value is
Medium.
8
Click
to save new calibration
data.
The calibration data is not user specific.
Changes in calibration will affect all users of
the ESPRIT software.
Summary of Contents for QUANTAX EDS
Page 1: ...QUANTAX EDS Energy dispersive X ray spectrometer for electron microscopy User Manual ...
Page 6: ......
Page 8: ...User Manual 2 Safety Information 8 ...
Page 12: ......
Page 14: ...User Manual 3 The QUANTAX EDS System 14 ...
Page 22: ......
Page 24: ......
Page 48: ......
Page 50: ...User Manual 5 Step by Step Guides 50 ...
Page 54: ...User Manual 5 Step by Step Guides 54 Fig 5 2 1 Spectrometer calibration ...
Page 56: ...User Manual 5 Step by Step Guides 56 Fig 5 3 1 Image capture ...
Page 58: ...User Manual 5 Step by Step Guides 58 Fig 5 4 1 Image extension ...
Page 60: ...User Manual 5 Step by Step Guides 60 Fig 5 5 1 Image drift correction ...
Page 62: ...User Manual 5 Step by Step Guides 62 Fig 5 6 1 Image processing ...
Page 67: ...5 8 Using the Spectrum Chart QUANTAX EDS 67 Fig 5 8 1 Features in the spectrum chart ...
Page 72: ...User Manual 5 Step by Step Guides 72 Fig 5 10 2 Beam current setting ...
Page 75: ...5 11 Standards Library QUANTAX EDS 75 Fig 5 11 2 Quantification dialog ...
Page 97: ...5 14 Object Analysis QUANTAX EDS 97 Fig 5 14 1 Object analysis ...
Page 108: ...User Manual 5 Step by Step Guides 108 Fig 5 19 1 Map processing ...
Page 111: ...5 20 Phase Analysis QUANTAX EDS 111 Fig 5 20 1 Performing phase analysis ...
Page 113: ...5 21 Creating a Report QUANTAX EDS 113 Fig 5 21 1 Creating a report ...
Page 116: ......