
5.1 Preparatory Steps
QUANTAX EDS
51
This chapter gives the analyst an overview of the most important tools of the ESPRIT
software, permitting a quick start to perform analyses with the QUANTAX EDS system.
The procedures described here assume standard (default) ESPRIT settings. They also
refer to the use of a scanning electron microscope (SEM) or a scanning transmission
electron microscope (STEM) or similar. Differences in the analysis of bulk samples
(mainly SEM and FIB) and electron transparent samples (TEM, STEM, sometimes SEM or
FIB) are pointed out.
5.1
Preparatory Steps
This section describes the actions needed prior to starting the analytical work with the QUANTAX
EDS system.
Step
Examples/hints
1
Prepare the sample and place it in the
microscope
e.g., embedding, polishing and carbon
coating, TEM sample preparation
2
Adjust high voltage and working distance
(SEM) or eucentric height (TEM) on the
microscope. Switch on the beam. Adjust
magnification.
Choose high voltage (HV) based on
overvoltage of elements to analyze.
3
Start ESPRIT and log in.
Refer to section 4.1.
4
Use the
icon in the bottom left
corner of the
Microscope
configurator to
open the
MICROSCOPE
CONFIGURATION
dialog and check the
settings.
WD, Magnification, HV values must match
values set on the microscope (see section
4.5.3). Device status indicator should be
green.
5
Use the
icon in the bottom left
corner of the
EDS
configurator to open
the
EDS CONFIGURATION
dialog and
a) Set
Pulse throughput
and
Maximum
energy
Normal operation mode
should be selected.
For
Pulse throughput
and
Maximum
energy
use
Automatic
settings. Change it on
demand.
b) Check
Detector temperature.
Check operation temperature. Device status
indicator should be green (see section 4.5.5).
6
Check input count rate and dead time for
the analysis.
The input count rate should be chosen
according to the analytical task.
7
Use the
icon in the bottom left
corner of the
Sample
configurator to
open the
SAMPLE PROPERTIES
dialog
and set sample name and description.
The sample name applies for all further
measurements.
8
Proceed with required analysis.
Refer to section 5.7, 5.14, 5.15 or 5.17.
Summary of Contents for QUANTAX EDS
Page 1: ...QUANTAX EDS Energy dispersive X ray spectrometer for electron microscopy User Manual ...
Page 6: ......
Page 8: ...User Manual 2 Safety Information 8 ...
Page 12: ......
Page 14: ...User Manual 3 The QUANTAX EDS System 14 ...
Page 22: ......
Page 24: ......
Page 48: ......
Page 50: ...User Manual 5 Step by Step Guides 50 ...
Page 54: ...User Manual 5 Step by Step Guides 54 Fig 5 2 1 Spectrometer calibration ...
Page 56: ...User Manual 5 Step by Step Guides 56 Fig 5 3 1 Image capture ...
Page 58: ...User Manual 5 Step by Step Guides 58 Fig 5 4 1 Image extension ...
Page 60: ...User Manual 5 Step by Step Guides 60 Fig 5 5 1 Image drift correction ...
Page 62: ...User Manual 5 Step by Step Guides 62 Fig 5 6 1 Image processing ...
Page 67: ...5 8 Using the Spectrum Chart QUANTAX EDS 67 Fig 5 8 1 Features in the spectrum chart ...
Page 72: ...User Manual 5 Step by Step Guides 72 Fig 5 10 2 Beam current setting ...
Page 75: ...5 11 Standards Library QUANTAX EDS 75 Fig 5 11 2 Quantification dialog ...
Page 97: ...5 14 Object Analysis QUANTAX EDS 97 Fig 5 14 1 Object analysis ...
Page 108: ...User Manual 5 Step by Step Guides 108 Fig 5 19 1 Map processing ...
Page 111: ...5 20 Phase Analysis QUANTAX EDS 111 Fig 5 20 1 Performing phase analysis ...
Page 113: ...5 21 Creating a Report QUANTAX EDS 113 Fig 5 21 1 Creating a report ...
Page 116: ......