
User Manual
5 Step-by-Step Guides
78
5.11.3
Maintain Standards Library for Electron Transparent Samples –
Zeta Factors
This guide describes how to set up the Zeta factor standards library. Quantification of electron
transparent samples with the Zeta factor method requires knowledge of the beam current in [pA]
for standard and “unknown” sample and sample thickness of the standard sample.
Experimental Zeta Factors
Follow this guide to update Zeta factors for the elements contained in the standard sample.
Step
Example/hints
1
Set up a Cliff-Lorimer standards library.
Follow step 1-4 in section 5.11.2 since the
geometry setup for Cliff-Lorimer and Zeta
factors is the same.
2
Select the
or the
workspace and
Acquire or open a spectrum of the
reference/standard sample. The composition
and thickness of the sample has to be known.
The beam current during measurement also
has to be known and has to be entered as
described in 5.13.3 or, if offline, as spectrum
parameter: double click on the highlighted line
of the active spectrum in the spectrum list
(below the spectrum) to open the
SPECTRUM PROPERTIES
dialog and enter
the beam current in pA as system factor in
the
Parameter
tab.
a)
a spectrum
b) Use the Spectrum chart
icon to
open a spectrum.
3
Follow the standardless quantification
steps 6 to 13 to identify elements, set
background regions and deconvolution.
Refer to section 5.12.2. Set quantification
method mode to
(interactive) to access
the
QUANTIFICATION
dialog.
4
Select Zeta factor method under
Quantification model.
5
Click
.
The
EDIT STANDARD PROPERTIES
dialog
pops up.
6
Enter reference element concentration
values and sample thickness for standard
sample.
Review Beam current. If missing or incorrect,
enter the correct value and also enter the
correct value in the spectrum
Parameter
tab
(see step 2).
The density will be calculated internally using
the sample composition, if changed manually,
the thickness will change, respectively.
7
Click
.
The
VALIDATION
dialog pops up with new
and old Zeta factors. Initial Zeta factors have
the same values as theoretical Cliff-Lorimer
factors (and are considerably smaller).
Summary of Contents for QUANTAX EDS
Page 1: ...QUANTAX EDS Energy dispersive X ray spectrometer for electron microscopy User Manual ...
Page 6: ......
Page 8: ...User Manual 2 Safety Information 8 ...
Page 12: ......
Page 14: ...User Manual 3 The QUANTAX EDS System 14 ...
Page 22: ......
Page 24: ......
Page 48: ......
Page 50: ...User Manual 5 Step by Step Guides 50 ...
Page 54: ...User Manual 5 Step by Step Guides 54 Fig 5 2 1 Spectrometer calibration ...
Page 56: ...User Manual 5 Step by Step Guides 56 Fig 5 3 1 Image capture ...
Page 58: ...User Manual 5 Step by Step Guides 58 Fig 5 4 1 Image extension ...
Page 60: ...User Manual 5 Step by Step Guides 60 Fig 5 5 1 Image drift correction ...
Page 62: ...User Manual 5 Step by Step Guides 62 Fig 5 6 1 Image processing ...
Page 67: ...5 8 Using the Spectrum Chart QUANTAX EDS 67 Fig 5 8 1 Features in the spectrum chart ...
Page 72: ...User Manual 5 Step by Step Guides 72 Fig 5 10 2 Beam current setting ...
Page 75: ...5 11 Standards Library QUANTAX EDS 75 Fig 5 11 2 Quantification dialog ...
Page 97: ...5 14 Object Analysis QUANTAX EDS 97 Fig 5 14 1 Object analysis ...
Page 108: ...User Manual 5 Step by Step Guides 108 Fig 5 19 1 Map processing ...
Page 111: ...5 20 Phase Analysis QUANTAX EDS 111 Fig 5 20 1 Performing phase analysis ...
Page 113: ...5 21 Creating a Report QUANTAX EDS 113 Fig 5 21 1 Creating a report ...
Page 116: ......