System Factor Calibration/ Beam Current Setting
For standard-based quantification, a probe current-dependent system factor has to be determined.
There are two ways to proceed depending on sample type and whether a picoamperemeter /
beam current monitor is available or not. The system factor value is stored in the ESPRIT software
and saved along with the following measurements until the software is closed. The quantification
of electron transparent samples with the Zeta-factor method requires the knowledge of
beam/probe current.
System Factor Calibration
Step
Example/hints
1
Move the calibration sample into the
analytical position and adjust microscope
parameters.
Use a single element calibration sample (Co,
Mn, Cu). Select a homogeneous area for
spectrum acquisition.
2
Load or create a standards library.
Follow section 5.11.1.
3
Click the
Standards configurator
’s
calibration icon
to open the
SYSTEM FACTOR CALIBRATION
dialog.
4
Select the material of the
Calibration
sample
from the
Select the sample to
use for calibration
drop-down list.
5
Set
Acquisition time
parameters.
6
Click
to calibrate the
system factor.
After the calibration is finished the message
System is now calibrated
appears. Do not
change microscope parameters afterwards.
7
the
SYSTEM FACTOR
CALIBRATION
dialog.
Spectra that are obtained after this calibration
will contain the measured system factor.
8
Proceed with measurement of standard
samples.
Continue with 5.12.3 for standard-based
measurement using a temporary reference.
Continue with 5.12.4 for standard-based
measurement using a standards library.
Summary of Contents for QUANTAX EDS
Page 1: ...QUANTAX EDS Energy dispersive X ray spectrometer for electron microscopy User Manual ...
Page 6: ......
Page 8: ...User Manual 2 Safety Information 8 ...
Page 12: ......
Page 14: ...User Manual 3 The QUANTAX EDS System 14 ...
Page 22: ......
Page 24: ......
Page 48: ......
Page 50: ...User Manual 5 Step by Step Guides 50 ...
Page 54: ...User Manual 5 Step by Step Guides 54 Fig 5 2 1 Spectrometer calibration ...
Page 56: ...User Manual 5 Step by Step Guides 56 Fig 5 3 1 Image capture ...
Page 58: ...User Manual 5 Step by Step Guides 58 Fig 5 4 1 Image extension ...
Page 60: ...User Manual 5 Step by Step Guides 60 Fig 5 5 1 Image drift correction ...
Page 62: ...User Manual 5 Step by Step Guides 62 Fig 5 6 1 Image processing ...
Page 67: ...5 8 Using the Spectrum Chart QUANTAX EDS 67 Fig 5 8 1 Features in the spectrum chart ...
Page 72: ...User Manual 5 Step by Step Guides 72 Fig 5 10 2 Beam current setting ...
Page 75: ...5 11 Standards Library QUANTAX EDS 75 Fig 5 11 2 Quantification dialog ...
Page 97: ...5 14 Object Analysis QUANTAX EDS 97 Fig 5 14 1 Object analysis ...
Page 108: ...User Manual 5 Step by Step Guides 108 Fig 5 19 1 Map processing ...
Page 111: ...5 20 Phase Analysis QUANTAX EDS 111 Fig 5 20 1 Performing phase analysis ...
Page 113: ...5 21 Creating a Report QUANTAX EDS 113 Fig 5 21 1 Creating a report ...
Page 116: ......