Weight
Standard Power Supply: .. .. .. .. .. 28.0 lbs. (12.7 kg.)
Siemens Power Supply: . .. .. .. .. .. 36.0 lbs. (16.3 kg.)
Control Panel and Stem: .. .. .. .. .. 7.5 lbs. (3.4 kg.)
Pedestal Cart: . .. .. .. .. .. .. .. .. .. .. 52.0 lbs. (23.6 kg.)
Injector Head: . .. .. .. .. .. .. .. .. .. .. 45.0 lbs. 20.4 kg.)
CVi Adjustable Arm: . .. .. .. .. .. .. .. 1.45 lbs (0.66 Kg.)
CVi Control Panel and Stem: .. .. .. 7.0 lbs. (3.18 Kg.)
Cord Lengths
Power cord: .. .. .. .. .. .. .. .. .. .. .. .. 12 feet (3.7 meters)
Control console:.. .. .. .. .. .. .. .. .. .. 10 feet (3 meters)
Injector Head power: .. .. .. .. .. .. .. 5 feet (1.5 meters)—mobile cart mounted
Specifi cations that Apply to UL Labeled Product
Classi
fi
cation
Type of protection against electric shock: .. .. .. .. Class 1
Degree of protection against electric shock: . .. .. Type CF applied part
Degree of protection against ingress of water: .. Ordinary
Methods of sterilization or disinfecting: .. .. .. .. .. None
Mode of operation: . .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. Continuous
Note: Equipment not suitable for use in the presence of a
fl
ammable anesthetic mixture with air or
with oxygen or nitrous oxide.
Transportation and Storage Requirements
Type of Temperature Range
Injection System
Patient Kits
Ambient temperature range
-40º to 122º F (-40º to 50º C)
0º to 122º F (-18º to 50º C)
Relative humidity range
10 to 95% non-condensing
Same
Atmospheric pressure range
7 to 15 psi (50 to 106 kPa)
10 to 15 psi (70 to 106 kPa)
Operating Environment Requirements
Type of Temperature Range
Injection System
Patient Kits
Ambient temperature range
18º C to 29.4º C (64º F to 85ºF) Same
Relative humidity range
10 to 95% non-condensing
Same
Atmospheric pressure range
10 to 15 psi (70 to 106 kPa)
Same
For UL labeled product. UL/c-UL Classi
fi
ed Mark, “Medical Electrical Equipment”, Control # 17ZM,
UL2601-1, CLASSIFIED BY UNDERWRITERS LABORATORIES INC.® WITH RESPECT TO
ELECTRIC SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH
UL 2601-1, and CAN/CSA C22.2 NO. 601.1.
W
eight, UL
Specs and Environmental Speci
fi cations
Speci
fi cations
P/N 900468-001 Rev. 02
ACIST Medical Systems, Inc.
Section 8 Speci
fi
cations
71
Summary of Contents for ACIST CVi CMS2000
Page 1: ...ACIST Contrast Delivery System User Manual 900468 001 02 2010 01 and CMS2000 and E2000Voyager...
Page 2: ...This page intentionally left blank...
Page 14: ...Information Warnings xii ACIST Medical Systems Inc P N 900468 001 Rev 02...
Page 91: ...P N 900468 001 Rev 02 ACIST Medical Systems Inc Notes 77 Notes...