System Setup
28
Section 3 System Setup
ACIST Medical Systems, Inc.
P/N 900468-001 Rev. 02
19. Attach the hand syringe to the manifold and slowly draw saline
through the tubing to make sure all air is removed from the lines,
the transducer cartridge, and the manifold.
Tip:
The “start, stop, and go” technique often works well for
this. Start drawing saline until about 6 inches of saline is
in the tubing and then stop. After a brief pause, continue
slowly drawing the saline until the entire tubing is full.
Remember, drawing at a slow rate will help avoid the
formation of air bubbles in the tube.
Attach the hand syringe to the
manifold and slowly draw saline
through the tubing
20. Open the saline pump door by raising the pump handle and
moving it counterclockwise from the right side of the pump to the
left side.
21. Place the tubing in the saline pump, leaving some slack on
either side of the pump.
22. Position the black v-teeth guides all the way to the bottom and
then move them up to the next highest locking position. Verify
that the v-teeth are centered over the tubing. The guides hold
the tubing in place and prevent kinking.
23. Close the saline pump door by moving the handle from the left
side to the right side to its locked position.
24. Remove the hand syringe and empty it into a waste container.
25. Press DONE on the touchscreen.
Place the saline tubing in the
saline pump, making sure the
black v-teeth guides are lowered
on each side and centered over
the tubing
Loading the
Angiographic Kit
18. Connect the spike from the saline tubing to the saline bag and
hang the saline bag on the hanger located on the right upper
corner of the injector head.
Important:
Make sure the saline bag is fully spiked. Otherwise,
air will be introduced into the tubing.
Spike and hang
saline bag
Summary of Contents for ACIST CVi CMS2000
Page 1: ...ACIST Contrast Delivery System User Manual 900468 001 02 2010 01 and CMS2000 and E2000Voyager...
Page 2: ...This page intentionally left blank...
Page 14: ...Information Warnings xii ACIST Medical Systems Inc P N 900468 001 Rev 02...
Page 91: ...P N 900468 001 Rev 02 ACIST Medical Systems Inc Notes 77 Notes...