Performing Patient Procedures
P/N 900468-001 Rev. 02 ACIST Medical Systems, Inc.
Section 5 Performing Procedures
49
Starting a New Case with the Existing Syringe
The CMS2000, Voyager, and CVi systems are designed to allow the
use of the multi-procedural syringe kit for up to
fi
ve (5) procedures
within 12 hour period. This allows you to save the contrast that
remains in the syringe at the end of a case.
Note:
The CL100H is not designed for re-use of the syringe kit
beyond a single procedure.
To end a case and start a new case using the same syringe:
1. Before ending the previous case, record all case information
(for example, CONTRAST DELIVERED, LAST INJECTION,
etc.).
This information is cleared after the case is ended.
2. Press END CASE.
3. Press
OK.
The following prompt is displayed: Shut Down System?
4. Press
NO.
The screen displays the following information:
Maximum recommended cases per syringe : 5
Number of cases this syringe: X
(X= number of cases counted by the system)
The system will recommend using a new syringe if the current
number of cases equals or exceeds the recommended number
of cases per syringe.
If you exceed the recommended number of cases, the
performance of the patient kit may be degraded.
The following prompt is displayed: Use new syringe?
5. Press
NO.
6. Close the white contrast slide clamp.
7. Disconnect and discard the catheter, high pressure (injection)
tubing and stopcock, and the automated manifold kit, including
the manifold, saline connection tubing to the syringe and high
pressure (injection) tubing.
Properly discard disposables in accordance with all local, state,
and federal regulations, codes and directives.
Note:
If the syringe will not be used immediately, place the pre-
sterile cap (provided in the automated manifold kit) on the
syringe and tighten it.
8. Connect the new automated manifold and hand controller kit
assemblies to the syringe.
9. Continue with the rest of the message prompts to start a new
case using the existing syringe.
Starting Another Case with a New Syringe
To end a case and start another case with a new syringe:
1. Before ending the previous case, record all case information
(for example, CONTRAST DELIVERED, LAST INJECTION,
etc.).
This information is cleared after the case is ended.
End Case? screen
Close the white contrast slide
clamp
Use new syringe? screen
Starting a New Case wtih the Existing Syringe
End Case? screen
Summary of Contents for ACIST CVi CMS2000
Page 1: ...ACIST Contrast Delivery System User Manual 900468 001 02 2010 01 and CMS2000 and E2000Voyager...
Page 2: ...This page intentionally left blank...
Page 14: ...Information Warnings xii ACIST Medical Systems Inc P N 900468 001 Rev 02...
Page 91: ...P N 900468 001 Rev 02 ACIST Medical Systems Inc Notes 77 Notes...