Injection Parameter Ranges Table
System
Injection type
FLOW
1
(ml/s)
VOLUME
2
(ml)
PRESSURE
3
Pounds per
square inch – psi
(Kilopascals–kPa)
RISE TIME
4
(Sec)
CL100H,
CMS2000, and
CVi (Cardiac
mode)
LCA
0.8 – 10.0
0.8 – 20.0
200 –1,200
(1,379 – 8,274)
0.0-1
RCA
0.8 – 10.0
0.8 – 20.0
LV/Ao
0.8 – 40.0
0.8 – 99.9
Other
0.8 – 40.0
0.8 – 99.9
Voyager and
CVi (Peripheral
mode)
Pigtail
0.8 – 40.0
0.8 – 99.9
200 –1,200
(1,379 – 8,274)
0.0–1
Selective
0.8 – 15.0
0.8 – 99.9
200 –1,200
(1,379 – 8,274)
Microcatheter
0.8 – 3.0
0.8 – 10.0
200 –300
(1,379 – 2,068)
Other
0.8 – 40.0
0.8 – 99.9
200 –1,200
(1,379 – 8,284)
Select Mode
•
INJECT: Arms the system, readying it to inject contrast.
•
SALINE: Enables you to
fl
ush saline through the system at a
rate of 1.66 ml per second. There is a 10 second timeout on
saline
fl
ushes.
•
PURGE: Enables you to purge either the syringe or the high-
pressure tubing.
- Purging the syringe clears the air from the syringe through
the contrast valve and into the contrast container.
- Purging the high pressure (injection) tubing requires
turning the 3-way stopcock off to the patient.
•
KVO (Keep Vein Open, or Keep Vessel Open): Available only on
Voyager and CVi (peripheral mode) models. Provides a periodic
pulse of saline to keep the
fl
uid pathway open to the patient
during delays between injections. This feature consists of two
screen controls.
To use KVO,
fi
rst specify the amount of saline to be dispensed
during each pulse by pressing the KVO Rate parameter. The
available range for this parameter is 0.1 – 10 ml/min. There is a
20-minute timeout on the KVO function, meaning a maximum of
200 ml of saline can be dispensed during any single activation.
Injection Parameter Ranges
Parameter ranges for each injection type are shown in the table below
Select Mode
Injection Parameters
1. In 0.1 ml/sec increments.
2. In 0.1 ml increments.
3. In 1 psi (1 kPa) increments.
4. In 0.1 sec. increments.
About the T
ouchscreen
36
Section 4 Touchscreen
ACIST Medical Systems, Inc.
P/N 900468-001 Rev. 02
Summary of Contents for ACIST CVi CMS2000
Page 1: ...ACIST Contrast Delivery System User Manual 900468 001 02 2010 01 and CMS2000 and E2000Voyager...
Page 2: ...This page intentionally left blank...
Page 14: ...Information Warnings xii ACIST Medical Systems Inc P N 900468 001 Rev 02...
Page 91: ...P N 900468 001 Rev 02 ACIST Medical Systems Inc Notes 77 Notes...