Message
Probable Cause
Corrective
Action
Report to
Service
Non-system disk or disk error
System is experiencing problem reading
the software chip.
Call a service representative.
NVRAM Failure!
The NVRAM failed a self-test during
startup.
Call a service representative.
Pressure Limit (Indicators box message)
The system is limiting the pressure by
decreasing the motor speed (
fl
ow
rate).
Because the system automatically limits the
pressure, the selected
fl
ow rate may not be
reached. Otherwise, you do not need to take
any corrective action unless you see the
“Pressure Limit Exceeded” message (see
below).
Pressure Limit (Indicators box message)
The system is limiting the pressure by
decreasing the motor speed (
fl
ow
rate).
Because the system automatically limits the
pressure, the selected
fl
ow rate may not be
reached. Otherwise, you do not need to take
any corrective action unless you see the
“Pressure Limit Exceeded” message (see
below).
Pressure Limit Exceeded (1)
Pressure Limit Exceeded (2)
NOTE: Use the corrective actions for
either of these messages.
The
notations (1) and (2) are useful to
report to a service representative,
if a call for service is needed.
•Stopcock is not open to allow injection.
•Catheter is kinked. •Preset pressure is
too low for required injection.
1. Make sure stopcock is open and catheter
is not kinked. 2. Increase the preset pressure.
3. Decrease the
fl
ow rate. 4. Increase the rise
time.
PURGE is Incomplete.
The front stop is reached before the
syringe valve closes.
The syringe purge
function removes air from the syringe.
When all air is removed, contrast will
force the syringe valve closed, triggering
a successful purge. If the purge fails to
close the syringe valve, this message is
displayed.
1. Do a re
fi
ll (press ST
AR
T
REFILL), and then
Purge. 2. Retry syringe purge.
After the purge
is complete, the system is ready to be armed
for the next series of injections, or 3. Replace
patient kit, or 4. Call a service representative
(syringe valve sensor may be defective).
Ram Forward (Indicators box message)
Ram has reached the forward position.
Informational only
.
Ram Reverse (Indicators box message)
Ram has reached the reverse position.
Informational only
.
T
roubleshooting
T
roubleshooting
P/N 900468-001 Rev. 02
ACIST Medical Systems, Inc.
Section 7 Troubleshooting
65
Summary of Contents for ACIST CVi CMS2000
Page 1: ...ACIST Contrast Delivery System User Manual 900468 001 02 2010 01 and CMS2000 and E2000Voyager...
Page 2: ...This page intentionally left blank...
Page 14: ...Information Warnings xii ACIST Medical Systems Inc P N 900468 001 Rev 02...
Page 91: ...P N 900468 001 Rev 02 ACIST Medical Systems Inc Notes 77 Notes...