System Setup
P/N 900468-001 Rev. 02
ACIST Medical Systems, Inc.
Section 3 System Setup
25
The Load Syringe Assembly Screen is displayed when the
system is ready to load the syringe assembly:
7. Open the angiographic patient kits
•
Tighten the contrast spike to the syringe assembly.
•
Set aside the package containing the hand controller,
high-pressure injection tubing and high-pressure 3-way
stopcock. Place these items in the sterile
fi
eld.
•
Take the syringe and automated manifold kit components
from their sterile pouches and remove all caps.
Load syringe assembly screen
Tighten contrast spike on syringe
assembly
Open the chamber by pulling
the white button
Insert the syringe in chamber
and close the chamber door
completely
8. Open the rotating chamber door by pulling on the white pin and
lowering the door.
9. Insert the syringe into the chamber and then close the chamber
door completely, ensuring that the chamber is locked (make
sure the white latch pin clicks into place).
The chamber door must be closed for the system to function.
10. On the Load Syringe Assembly screen, press DONE.
Loading the
Angiographic Kit
Load syringe assembly screen