System Maintenance
P/N 900468-001 Rev. 02
ACIST Medical Systems, Inc.
Section 6 Maintenance
53
Cleaning the Injector Head
1. Make sure system power is turned off.
2. Clean the entire injector head using a soft cloth moistened with
warm water.
CAUTION: Be careful not to get the injector head so wet that
water drips inside. Do not immerse the injector
head in water. Do not spray or pour liquid directly
onto the display panel. Do not use excessive
pressure.
Pressure Transducer Cartridge and Backplate
The disposable pressure transducer cartridge and the backplate
it mounts to both have rubber membranes. When cleaning these
components you can wipe them with a damp cloth, but be careful
not to soak these components or expose them to bleach or other
cleaning agents. If the backplate’s rubber membrane looks white
or dull you can try applying a small amount of mineral oil. If a good
pressure waveform cannot be achieved, one possible cause may
be a worn or damaged rubber membrane.
Important!
Do not soak the rubber membrane or expose it to
bleach or other cleaning agents.
Decontamination
If the system needs cleaning to remove any biohazardous
materials (blood, etc.), the following are acceptable solutions to
use.
•
A 12% bleach solution (only on plastic components)
•
Hospital grade disinfectant
•
An alcohol pad
Wear protective gloves. Apply the solution to a soft cloth and wipe
down the system. Discard all materials used to decontaminate the
product in accordance with all local, state, and federal regulations,
codes and directives.
CAUTION:
Do not apply these solutions to the rubber
membrane of the pressure transducer backplate
because this may cause the transducer to degrade,
resulting in poor pressure waveforms.
Use a soft cloth moistened with
warm water to clean the injector
head
Clean the transducer backplate
membrane with a damp cloth
Cleaning the Injector Head
Summary of Contents for ACIST CVi CMS2000
Page 1: ...ACIST Contrast Delivery System User Manual 900468 001 02 2010 01 and CMS2000 and E2000Voyager...
Page 2: ...This page intentionally left blank...
Page 14: ...Information Warnings xii ACIST Medical Systems Inc P N 900468 001 Rev 02...
Page 91: ...P N 900468 001 Rev 02 ACIST Medical Systems Inc Notes 77 Notes...