System Setup
24
Section 3 System Setup
ACIST Medical Systems, Inc.
P/N 900468-001 Rev. 02
When the calibration is complete the language selection screen
is displayed.
Language Selection Screen
4. To switch to a different language, press the appropriate
language button within 5 seconds.
The software defaults to the last selected language if none is
selected in 5 seconds.
(Optional) If you select a language other than US English, the
Pressure Selection screen is displayed. You can select the unit
of measure used for pressure: Pounds per square inch (psi)
or Kilopascals (kPa). After you select psi or kPa, this unit of
measure will continue to display until you change it. The system
defaults to the last selected pressure unit if none is selected
within 5 seconds.
The Start screen is displayed.
5. Press START to continue with system setup.
Note:
Do not place the syringe assembly into the chamber until
the system completes the calibration process.
The Cardiac/Peripheral selection screen is displayed:
6. Select the type of procedure you wish to perform.
Your selection dictates the set of pre-de
fi
ned paramater
selections that will be loaded into the machine.
•
Cardiac: This option is best suited for heart-related
procedures. The following injection types will be made
available: LCA, RCA, LV/Ao, and Other.
X-ray functions are not available.
•
Peripheral: This option is best suited for procedures not
related to the heart (for example: arms, legs, kidneys,
carotid artery, etc). The following injection types will be
made available: Pigtail, Selective, Microcatheter, and Other.
X-ray system synchonization is available, provided the
proper X-ray interface cable is installed.
After selecting the procedure type, ram calibration will begin.
After ram calibration the “Ready to load syringe” screen is
displayed.
Language selection screen
Start screen
CVi Cardiac or Peripheral
selection screen
CVi Ready to load syringe screen
Positioning Ram to hold syringe
assembly
Setup Screens
Pressure selection screen
Summary of Contents for ACIST CVi CMS2000
Page 1: ...ACIST Contrast Delivery System User Manual 900468 001 02 2010 01 and CMS2000 and E2000Voyager...
Page 2: ...This page intentionally left blank...
Page 14: ...Information Warnings xii ACIST Medical Systems Inc P N 900468 001 Rev 02...
Page 91: ...P N 900468 001 Rev 02 ACIST Medical Systems Inc Notes 77 Notes...