Message
Probable Cause
Corrective
Action
Report to
Service
Chamber Not Closed (Status window
message)
Chamber Open (Indicators box message)
Chamber is open.
The chamber must be
fully closed for proper operation of the
system. Injector motor moves (inject,
syringe purge, tubing purge and re
fi
ll)
will
not be performed if the chamber is not
closed.
A
magnetic switch located in the
chamber door triggers the sensor
.
1. Rotate the chamber to the closed position.
Make sure the white latch pin clicks into place.
2. If the door is closed and the problem
persists, call a service representative.
Check Manifold V
alve. Remove syringe
assembly from chamber …OR Check the
manifold valve sensor
.
This message displays during setup if the
system detects a patient kit in place before
system calibration.
Remove the syringe assembly from the
chamber or check the Manifold V
alve Sensor
.
Check Syringe V
alve. Remove syringe
assembly from chamber …OR Clean
syringe valve sensor and check cable
connection.
This message displays during setup if the
system detects a patient kit in place before
system calibration.
Remove the syringe assembly from the
chamber or check the Syringe V
alve Sensor
.
Communication Error Has Occurred.
Communication T
imeout!
Communications errors.
1. Shut down the system and power up again.
2. Call a service representative to report this
message.
√
Contrast Empty (Indicators box message)
No contrast has been detected by
the Contrast Sensor
. This message
is displayed in the Indicators Box (the
message “No Contrast” is displayed in the
Status Window).
1. Make sure the contrast spike tubing is in the
Contrast Sensor
.
2. Make sure there is contrast in the container
,
or connect a new container
.
3. Make sure the Contrast Sensor (the black
sensor that clamps around the contrast
spike) is clean.
4. Make sure the Contrast Sensor is closed
tightly
.
5. Make sure the white contrast spike clamp is
open.
6. Make sure there are no air bubbles trapped
in the spike.
7. Call a service representative.
Counter/Position Mismatch.
During an injection, the system monitors
two separate components to determine
location of the ram in the syringe.
This
message will be displayed if these two
separate feedback loops do not agree.
Press
OK
to continue. If the problem persists,
call a service representative.
T
roubleshooting
60
Section 7 Troubleshooting
ACIST Medical Systems, Inc.
P/N 900468-001 Rev. 02
Summary of Contents for ACIST CVi CMS2000
Page 1: ...ACIST Contrast Delivery System User Manual 900468 001 02 2010 01 and CMS2000 and E2000Voyager...
Page 2: ...This page intentionally left blank...
Page 14: ...Information Warnings xii ACIST Medical Systems Inc P N 900468 001 Rev 02...
Page 91: ...P N 900468 001 Rev 02 ACIST Medical Systems Inc Notes 77 Notes...