Message
Probable Cause
Corrective
Action
Report to
Service
Standby!
The screen is locked and “Standby”
is displayed. If the Standby button
is depressed, all system activity is
suspended.
1. Check the cable connections between the
control panel and the injector head.
2.
Assess your situation in the procedure.
If you are prepared to continue with the
procedure, release the Standby button
(located on right side of the control panel)
and press OK to continue.
Start Injection
V
oyager and CV
i only
Indicators box message.
This message will
display when the injection is initiated from
the imaging system, even if the “Inject
Delay” setting delays the actual start of the
injection.
Informational only
.
Syringe Not Detected
An installed syringe was not detected,
or the small metallic ball in the top of the
syringe is stuck in the up position.
This
error message will occur if you try to go
past the
fi
rst step in the setup process
without inserting the syringe into the
syringe chamber
.
1. Install a syringe. Load syringe into chamber
when screen displays “Load Syringe” (but
not before this).
2. Clean the Syringe V
alve Sensor
.
3. Use a new patient kit.
4. Call a service representative (sensor may
require replacing).
Syringe V
alve Closed
The small metallic ball in the top of the
syringe is stuck in the up position, or the
Syringe V
alve Sensor needs cleaning.
1. Do a re
fi
ll to pull the ball down.
2. Clean the sensor
. Make sure the gray wire
connection for the Syringe V
alve Sensor is
clean and plugged in, or
3. If the problem persists, change the
disposable syringe.
4. Call a service representative.
T
roubleshooting
P/N 900468-001 Rev. 02
ACIST Medical Systems, Inc.
Section 7 Troubleshooting
67
Summary of Contents for ACIST CVi CMS2000
Page 1: ...ACIST Contrast Delivery System User Manual 900468 001 02 2010 01 and CMS2000 and E2000Voyager...
Page 2: ...This page intentionally left blank...
Page 14: ...Information Warnings xii ACIST Medical Systems Inc P N 900468 001 Rev 02...
Page 91: ...P N 900468 001 Rev 02 ACIST Medical Systems Inc Notes 77 Notes...