System Setup
30
Section 3 System Setup
ACIST Medical Systems, Inc.
P/N 900468-001 Rev. 02
Important: The remaining steps of the setup require a tech or operator to be scrubbed in.
The AngioTouch Hand Controller Setup screen is displayed:
32. Apply a sterile drape over the touchscreen.
This enables the operator or scrub tech to continue the screen
setup in the next section.
33. Empty the contents of the hand controller package on the sterile
fi
eld.
This package contains the pneumatic AngioTouch hand
controller tubing, high-pressure(injection) tubing, and high-
pressure 3-way stopcock.
34. The scrub person should hand one end of the high-pressure
(injection) tubing to someone outside the sterile
fi
eld. That
person should then attach it to the manifold tubing and then
place the tubing through the air column sensor. At the same time
the scrub person can connect the 3-way stopcock to the other
end of the tubing.
Note:
The air column detect sensor is not a substitute for
user vigilance.
AngioTouch Hand Controller
setup screen
High pressure (injection) tubing
placed in air column detect
sensor
35. The scrub person should hand off the two connectors from
the end of the pneumatic hand controller so that they can be
connected to the control panel.
The two lines are connected to the bottom right side of the
control panel. The control panel can be tipped back for easier
access to the connectors.
Important:
Make sure the connections are screwed on straight
and tight. Do not over-tighten.
Connect the hand controller to
the control panel
Saline
connector
Contrast
connector
The person who is scrubbed in will complete the following steps. Make sure hand controller
connections are tight.
34. Press Done.
A screen similar to the Main Screen shown at the left is
displayed:
Main Screen
Status Window
The black section in the middle of the screen is the Status
Window, and is used to display system messages. Interactive
messages such as the calibration message shown here are
displayed in this section. These messages require a response.
Loading the
Angiographic Kit
Summary of Contents for ACIST CVi CMS2000
Page 1: ...ACIST Contrast Delivery System User Manual 900468 001 02 2010 01 and CMS2000 and E2000Voyager...
Page 2: ...This page intentionally left blank...
Page 14: ...Information Warnings xii ACIST Medical Systems Inc P N 900468 001 Rev 02...
Page 91: ...P N 900468 001 Rev 02 ACIST Medical Systems Inc Notes 77 Notes...