Message
Probable Cause
Corrective
Action
Report to
Service
Hand Controller Calibration Failed.
Hand Controller requires Calibration.
These error messages can occur if:
1.
Y
ou press the Contrast Button on the
hand controller before the message
prompts you to do this.
2.
Y
ou accidentally press the Saline Button
instead of the Contrast Button.
3.
The hand controller was not calibrated
at setup.
4. Connection to the touchscreen is not
tight.
5. Defective hand controller
.
1. Make sure you wait until the message
displays on screen to press the Contrast
Button (do not press the Saline Button).
2. Check to see if the hand controller
attachments are secure (at bottom of the
touchscreen), or
3.
To
retry hand controller calibration, press
the CALIBRA
TE HC key
.
NOTE:
The hand controller must be calibrated
successfully each time you begin a case.
If you disconnect or change a hand
controller at any time during a case, you
must calibrate it.
Injection Motor Malfunction (OS).
Injection Motor Malfunction (OT).
1. Internal error
.
2. For OT
errors, there may be a restriction
in the
fl
uid
fl
ow
.
1. Make sure the stopcock is open and that
fl
uid
fl
ow is not restricted in any part of the
system or patient kit.
2. W
rite down the screen message (and
whether it says “OS” or “OT”), and call
a service representative to report this
message.
√
Interprocessor Ping Error
.
Communications error
.
1.
T
urn system power of
f and on again.
2. If the problem persists call a service
representative to report this message.
√
Invalid calibration data.
During initial startup calibration, the
system checks the calibration data.
If an error is detected, this message
is displayed. In addition to the error
message, an error description, error value,
and expected value will be displayed.
1. Press OK to retry calibration.
2. W
rite down the error message displayed on
screen and call a service representative to
report it.
√
T
roubleshooting
62
Section 7 Troubleshooting
ACIST Medical Systems, Inc.
P/N 900468-001 Rev. 02
Summary of Contents for ACIST CVi CMS2000
Page 1: ...ACIST Contrast Delivery System User Manual 900468 001 02 2010 01 and CMS2000 and E2000Voyager...
Page 2: ...This page intentionally left blank...
Page 14: ...Information Warnings xii ACIST Medical Systems Inc P N 900468 001 Rev 02...
Page 91: ...P N 900468 001 Rev 02 ACIST Medical Systems Inc Notes 77 Notes...