Injection Parameter Ranges
System
Injection type
FLOW
1
(ml/s)
VOLUME
2
(ml)
PRESSURE
3
Pounds per
square inch – psi
(Kilopascals–kPa)
RISE TIME
4
(sec)
CL100H,
CMS2000, and
CVi (Cardiac
mode)
LCA
0.8 – 10.0
0.8 – 20.0
200 –1,200
(1,379 – 8,274)
0.0-1
RCA
0.8 – 10.0
0.8 – 20.0
LV/Ao
0.8 – 40.0
0.8 – 99.9
Other
0.8 – 40.0
0.8 – 99.9
Voyager and
CVi (Peripheral
mode)
Pigtail
0.8 – 40.0
0.8 – 99.9
200 –1,200
(1,379 – 8,274)
0.0–1
Selective
0.8 – 15.0
0.8 – 99.9
200 –1,200
(1,379 – 8,274)
Microcatheter
0.8 – 3.0
0.8 – 10.0
200 –300
(1,379 – 2,068)
Other
0.8 – 40.0
0.8 – 99.9
200 –1,200
(1,379 – 8,284)
1. In 0.1 ml/sec increments.
2. In 0.1 ml increments.
3. In 1 psi (1 kPa) increments.
4. In 0.1 sec. increments.
Saline Rate
Automated
fi
xed
fl
ow rate of 100 ml/min.
Status Readouts
Contrast Remaining: Real time, total volume in Injection Syringe
Contrast Delivered: Real time, cumulative volume of contrast delivered per case in ml; reset with
each new case. Also reset if power is turned off and on again.
Last Injection:
0-99.9 ml @ 0-40 ml/sec (real time, peak/injection; reset each injection)
Program Control
• In
fi
nite single level injections
•
Variables entered through keypad on display screen for
fi
xed rate injections
Height
User-con
fi
gurable positioning system.
May be used as a cart and/or patient table (bed)-mounted system with remote control console.
Injection Parameters
Speci
fi cations
70
Section 8 Speci
fi
cations
ACIST Medical Systems, Inc.
P/N 900468-001 Rev. 02
Summary of Contents for ACIST CVi CMS2000
Page 1: ...ACIST Contrast Delivery System User Manual 900468 001 02 2010 01 and CMS2000 and E2000Voyager...
Page 2: ...This page intentionally left blank...
Page 14: ...Information Warnings xii ACIST Medical Systems Inc P N 900468 001 Rev 02...
Page 91: ...P N 900468 001 Rev 02 ACIST Medical Systems Inc Notes 77 Notes...