32
Instructions for use, ERGO 300 - BOWA- IFU-12930- ERG0300-S0-EN-20210701
6.5
Disassembly of the motor
Disassembling motor (three steps)
Disassemble motor before decontamination.
1. Unscrew handpiece holder from motor
2. Unscrew motor cap from motor
3. Pull motor plug from motor.
6.6
Disassembly of gear unit and Morcellator with protective tube
Disassembling seal unit
Remove membrane screw from seal unit and remove
membrane seal.
Remove seal retainer from gear unit and remove roof
shaped seal.
Risk of injury for the user!
The distal end of the cutting tube is very sharp, handle the cutting tube very carefully.
Removing protective tube and cutting tube
Make sure that the lock nut of the protective tube covers the
distal end of the cutting tube (protected
“NO CUT”
position
is visible).
Hold the gear unit with one hand so the distal end of the cut-
ting tube does not point to the user or any other person.
With the other hand carefully disconnect the protective tube
from the gear unit and pull it off.
Hold the cutting tube with the same hand and press it lightly
against the gear unit until it releases itself from the holder
and slides out smoothly. It can now be pulled out of the proxi-
mal opening of the gear unit.
1. Remove pro-
tective tube
2. Pull out cutting
tube