8. Inserting Obturator
Insert the Obturator through the seal unit, gear unit and
Trocar Cannula.
9. Attaching handle
Lift and hold the release lever (1) of the handle (2).
Push the handle (2) over the motor connection of the gear unit.
Release the handle release lever. The handle release lever must lock in position.
10. Connecting motor
Insert the motor coupling into the bottom end of the
handle.
Continue to insert the motor coupling until it locks into
the gear unit motor connection.
Align the red dot on the device connection plug of the
motor with the red dot on the motor connection of the
control unit.
Plug the motor device connection plug into the motor
connection of the control unit. The plug must lock in
position.
11. Connecting the pedal
Align the red dot on the device connection plug of the
pedal with the red dot on the pedal connection of the
control unit.
Plug the pedal device connection plug into the pedal con-
nection of the control unit. The plug must lock in position.