18
Instructions for use, ERGO 300 - BOWA- IFU-12930- ERG0300-S0-EN-20210701
Functional check
• Check Morcellator for loose parts
• Check that Morcellator is correctly assembled
• Are all labels easily readable?
• Check cutting tube: cutting tube evenly ground, no broken edges or bends?
• Switch on control unit
• Switch on motor (press pedal).
• Run motor with Morcellator for about 20 seconds; it must reach the set speed.
• If it does not run evenly, switch off motor with Morcellator.
• Switch off control unit.
The control unit will only operate if the pedal has been connected to the control unit.
• Check that Morcellator is correctly assembled.
• Check that the gear unit runs smoothly; if necessary treat it with NouvaOil care spray.
• Check that the motor is clean and runs smoothly; if necessary clean it and treat with NouvaOil care spray.
• Repeat the function check.
Risk of injury by motor failure
Problem: the motor becomes very hot. Cause: the motor was damaged during preparation.
Remedy: do not continue to use the motor. Always have a spare motor available.
5.3.2 Assembly of the Morcellator set with Trocar Cannula (according to
option 2
)
Using the Trocar Cannula
The Trocar Cannula is used instead of the protective tube. The integrated valve flap prevents the escape of gas when
instruments are moved in and out.
The nose of the Trocar Cannula partially extends beyond the cutter tube to protect organs in the abdominal cavity from
the sharp blade. This requires the Morcellator to be rotated under visual supervision (endoscopy) so the nose of the Tro-
car Cannula covers the cutter tube blade to protect the organ.
1. Visual check
• Check the product for sealing, corrosion and sterility.
• Ensure that it is free from rust, dents and scratches.
• The distal end of the Obturator must not be damaged.
• The sealing flap and the seal must not be damaged.
Risk of injury
Select blade. Make sure that the cutting edge of the blade is sharp and undamaged (e.g. no cracks or burrs).