Instructions for use, ERGO 300 - BOWA- IFU-12930- ERG0300-S0-EN-20210701 15
5.3.1 Mounting the morcellator set with protective tube (according to
option 1
)
There is a risk of injury to tissue by the sharp edge at the distal end of the cutting tube.
• Handle the cutting tube carefully.
• If the cutting tube is not used, rotate the protective tube to the
NO CUT
position.
• If the cutting tube is not used, it must be placed on a flat, sterile surface so it cannot fall and cause injury
• Check the cutting tube.
• Make sure that the cutting edge of the blade is sharp and undamaged (e.g. no cracks, deformation or burrs).
• Use only cutting tubes in perfect condition.
1. Inserting the cutting tube through the gear unit
Remove the seal unit from the gear unit.
Insert the cutting tube through the gear unit. The two
guide pins on the cutting tube must fit into the slot on
the flange of the gear unit.
2. Fixing the cutting tube to the gear unit
Hold the cutting tube with one hand.
Connect the sealing unit to the gear unit with the
bayo-net closure with the other hand.
3. Mounting protective tube
Select a suitable protective tube.
Push the protective tube over the distal end of the cutting
tube. The guide pin of the protective tube must be in the
distal slot of the gear unit.
Lock the protective tube in position with a click.