Prior to use
Make sure that the product has been correctly prepared and inspected.
Risk of injury
The ERGO 300 - Morcellator - control unit has been designed for operation with the morcellation blades as a set. The user
is solely responsible if different products are used with the set. Incompatibility may risk injury to patients.
Switching on and off
Switch on the control unit. The display lights up. A short signal tone sounds.
The control unit displays the value «200 rpm».
Operating mode (motor)
The motor is designed for intermittent operation (1 min. ON / 1 min. OFF, for 8 cycles).
6.1
Using the control unit
Setting the speed
The speed can be set manually in stages (within a range of 50 to 1000 rpm).
Press
to increase the speed.
Press
to reduce the speed.
The set speed is shown on the display.
Starting motor
Press pedal.
The blade rotates at the set speed.
The motor LED lights up.
To stop the motor remove foot from pedal.
Risk of injury and damage due to heat!
The Morcellator may generate heat if operated for a long period.
Remedy:
Use Morcellator only in intermittent mode of
«1 min. ON / 1 min. OFF, for 8 cycles»
.
Stop the Morcellator immediately if one or more components of the Morcellator set become very warm.
Precaution when changing parameters. Unusual behavior of instruments while operating may provoke false reactions and
jeopardize the patient.
Every setting and the new behavior of the instrument has to be verified and tested.