CUT
Instructions for use, ERGO 300 - BOWA- IFU-12930- ERG0300-S0-EN-20210701 27
6.2.2
Option 2
: Morcellation with Trocar Cannula
1. Hysterectomy
Select the procedure for the hysterectomy in accordance
with current medical standards.
2. Inserting Trocar Cannula
Insert Trocar Cannula with Obturator into an existing
access under visual control (endoscopy).
Remove Obturator.
3. Insert the cutting tube through the Trocar Cannula.
Actuate the valve flap lever of the Trocar Cannula and hold
in this position.
Carefully insert prepared cutting tube with gear unit and
seal unit into the proximal opening of the Trocar Cannula.
Cutting tube must be fully inserted.
Release the valve flap lever of the Trocar Cannula.
The guide pin of the Trocar Cannula must be in the distal
slot of the gear unit.
Lock the Trocar Cannula in position with a click.
Make sure that the cutting tube remains in the
“NO CUT”
position.
4. Inserting gripper
Insert gripper carefully through the proximal opening
of the gear unit and Trocar Cannula.
5. Preparing cutting tube
Rotate the Trocar Cannula so the
“CUT”
position is visible
from above.
If the cutting tube is completely inserted, the cut-ting edge
is now only partially covered by the Trocar Cannula.
Endoscope
Endoscope
Endoscope