6. Protecting the tissue
The nose on the Trocar Cannula, which partially covers the
sharp edge of the blade, must protect tissue that must not
be injured.
Rotate the Morcellator to the desired position for this.
7. Removing tissue
Grip the required tissue with the gripper.
The handle lock is set.
Start the Morcellator set motor.
Pull the tissue towards the rotating blade with the gripper.
Pull the gripper out of the patient through the gear unit.
The retracted mouth of the gripper must not touch the
blade.
Repeat the process until the morcellated tissue has been
completely removed.
Removing tissue without the use of the Gear Unit (
according to Option 2
)
The trocar sleeve is very well suited to remove further tissue from the abdomen via percutaneous access. To do this, place
the seal adapter on the trocar sleeve and screw the sealing unit onto the seal adapter.
Insert the grasping forceps under arthroscopic observation
through the sealing unit and the opened trocar sleeve into
the abdominal cavity.
Grasp the tissue with the grasping forceps and pull it out
through the trocar sleeve and sealing unit.
Protected area
Unprotected
cutting area
Endoscope
Endoscope