6. Inserting the Obturator
Insert the Obturator through the seal unit, gear unit and
protective tube to the stop.
7. Attaching the handle
Lift and hold the release lever (1) of the handle (2).
Push the handle (2) over the motor connection of the gear
unit.
Release the handle release lever. The handle release lever
must lock in position.
Do not open the handle release lever to remove the handle during the procedure.
8. Connecting the motor
Insert the motor coupling into the bottom end of the
handle.
Continue to insert the motor until it locks into the gear
unit’s motor connection.
Align the red dot on the device connection plug of the
motor with the red dot on the motor connection of the
control unit.
Plug the motor device connection plug into the motor
connection of the control unit. The plug must lock in
position.
9. Connecting pedal
Align the red dot on the device connection plug of the
pedal with the red dot on the pedal connection of the
control unit.
Plug the pedal device connection plug into the pedal
connection of the control unit.
The plug must lock in position.