7.1
Boreal Laser Inc.
GasFinderMC Operation Manual — Troubleshooting
July 17th 2017
GasFinderMC not operating, display not illuminated.
•
Check power supply.
- Plugged in? Is the supply socket live?
- Power cable intact?
Check the fuse.
There is a 3A fuse located at the left on
the rear panel.
The display shows “Low Light - No Reading”.
•
Check that the optical path is clear
.
- Vehicle/personnel obstruction? – move!
- Heavy fog? – wait.
- Heavy rain? – wait.
- Cobwebs/birds? – Clean and prevent a recurrence.
- Window dirty? – Clean (see page 6.2).
Verify optical alignment (see page 3.6).
Note
: If the system enters the “Low Light - No Reading” status
condition due to a temporary path obstruction, normal operation
will automatically resume once the obstruction has been removed
(e.g. fog lifts, rain stops or vehicle/equipment is moved).
No data transmission from GasFinderMC to the computer.
•
Check that the correct
baud rate
is selected (usually 9600).
-
Enter
System Menus
, and scroll to the
Serial Baud Rate
window.
Verify that the data cable is connected and the cable is OK (see page 3.11)
It is possible with a digital voltmeter to see a very short signal pulse
between pins 2 and 5 on the DB9 connector.
The light is always at the maximum of 16368.
•
To lower the light, remove or cover some of
the retrorefl ectors or position the refl ective
tape at an angle to the beam.
Note:
If there is too much light returning to the photo diode, it is
possible to damage the optical components. See Section 3.
Section 7. Troubleshooting
Symptom:
Solution:
Low Light
No Reading
Light Value Distance
16368
100m
Symptom:
Solution:
Symptom:
Solution:
Symptom:
Solution:
GasFinderMC
Fuse
Summary of Contents for GasFinderMC-2
Page 8: ...viii Boreal Laser Inc GasFinderMC Operation Manual Contents July 17th 2017 ...
Page 24: ...3 14 Boreal Laser Inc GasFinderMC Operation Manual Installation July 17th 2017 ...
Page 30: ...4 6 Boreal Laser Inc GasFinderMC Operation Manual Data I O July 17th 2017 ...
Page 71: ...BOREAL LASER INC MC Instrument Overall Dimensions JB0125MC A 1 of 1 NTS A Dim mm ...
Page 73: ......
Page 75: ......
Page 79: ......
Page 81: ......
Page 83: ......
Page 85: ......
Page 87: ......
Page 89: ......