GasFinderMC Operation Manual — Maintenance
July 17th 2017
5. System Stability and Data Quality Assurance
The
GasFinderMC
internal operating conditions may drift slightly with age and external
conditions. This is normal and the system will self adjust. Eventually the internal
adjustment will exceed a limit and a manual reset has to be applied. Information is
available in the $GFDBG string as the ‘centreline_duty_cycle’ and may be checked
periodically with the
GasViewMC
logging program. These should be monitored
regularly to check for system drifts outside normal operating ranges.
System performance indicators should be checked at least once per month. These
are best checked by connecting the
GasFinderMC
output to a computer with the
GasViewMC
program and collecting data for a minimum of one hour. A full day of
logging is recommended. The following system indicators can be read directly or
displayed from the log fi le using Excel macros from Boreal Laser Inc. These indicators
are helpful for quality assurance, system troubleshooting, trending and early detection
of potential problems to ensure system performance and reliability.
6. Performance Indicators
Array Dump
It is recommended that an array dump be performed monthly and the array fi le kept
for records as a history of system performance. The array dump shows the spectral
scan of the sample path and reference cell. This can be examined for changes against
previous fi les and saved for reference.
7. Annual System Check
A thorough system check and review of the following items should be made annually.
a. Inspect the system hardware for damage and integrity.
b. Check monthly performance parameters and review with historic data for system
performance and trending.
c. Inspect light levels throughout the system and make adjustments as required.
d. Check the condition of optical surfaces for damage or accumulation of foreign material.
e. Check and verify system output performance, signal levels and alarm performance
8. Factory Calibration and Maintenance
The
GasFinderMC
should be returned to the factory for calibration and service every
two years. This can be scheduled around maintenance activities to prevent associated
unit downtime.
Parameter
Normal
Low
High
centreline_duty_cycle
512
300
800
DAC
2000 - 3000
1800
4095
Reference cell quality
0.99
0 - 20%
n/a
Reference R2
0.99
0.90
n/a
Summary of Contents for GasFinderMC-2
Page 8: ...viii Boreal Laser Inc GasFinderMC Operation Manual Contents July 17th 2017 ...
Page 24: ...3 14 Boreal Laser Inc GasFinderMC Operation Manual Installation July 17th 2017 ...
Page 30: ...4 6 Boreal Laser Inc GasFinderMC Operation Manual Data I O July 17th 2017 ...
Page 71: ...BOREAL LASER INC MC Instrument Overall Dimensions JB0125MC A 1 of 1 NTS A Dim mm ...
Page 73: ......
Page 75: ......
Page 79: ......
Page 81: ......
Page 83: ......
Page 85: ......
Page 87: ......
Page 89: ......