9.3
Boreal Laser Inc.
GasFinderMC Installation Manual — Appendices
July 17th 2017
Appendix B — Status Codes
n
15
-
CAL_ERR
:
calibration line centre error
n
14
-
WAVG_ENAB
:
averaging enabled
n
13
-
NOT_CALIB
:
calibration not verifi ed
n
12
-
LOW_BAT
:
low battery indication
n
11
-
MENU_MODE
:
unit is currently in menu mode
n
10
-
CALIB_ERR
:
reference cell out of factory calibration
n
9
-
OVERFLOW
:
gas concentration >8 digits
n
8
-
OVERRANGE
:
receiver signal beyond gain switch control
n
7
-
LD_COLD:
TE Failure, laser diode off line
n
6
-
Lt_Lv_ERR
:
turbulence, timed out
n
5
-
NOLOCKDET
:
receiver unlocked, low signal
n
4
-
LOW_LIGHT
:
laser path blocked, low light
n
3
-
R2_LOW
:
R
2
lower than set minimum value
n
2
-
NOT_CTRD
:
laser not centred on gas wavelength
n
1
-
GAS_HIGH
:
concentration higher than set maximum
n
0
-
NO_ERROR
:
normal operation, no errors
Examples:
(a).
The above sample status code of “2800” corresponds to “0010_1000_0000_0000”. This indicates an instrument
status of:
n
13
- NOT_CALIB, n
11
- MENU_MODE
(b).
Code “A42E” would correspond to “1010_0100_0010_1110” and indicate an instrument status of:
n
15
-CAL_ERR,n
13
-NOT_CALIB,n
10
-CALIB_ERR,n
5
-NOLOCKDET,n
3
-R2_LOW,n
2
-NOT_CTRD,n
1
-GAS_HIGH
Under normall operating conditions, the most common status code is “1”. This is actually 0001, but the software ignores
the leading zeros and indicates a status code of “1”. This corresponds to “0000_0000_0000_0001”, and indicates an
instrument status of:
n
0
- NO_ERROR.
When the
GasFinderMC
is used in conjunction with ‘Hyperterminal’ or Boreal Laser’s
‘
GasViewMC
’ data-logging program, the data stream which is sent to the fi le is as follows:
Status Code 2 8 0 0
Hex n
3
n
2
n
1
n
0
0 0 0 0 0
1 0 0 0 1
2 0 0 1 0
3 0 0 1 1
4 0 1 0 0
5 0 1 0 1
6 0 1 1 0
7 0 1 1 1
8 1 0 0 0
9 1 0 0 1
A 1 0 1 0
B 1 0 1 1
C 1 1 0 0
D 1 1 0 1
E 1 1 1 0
F 1 1 1 1
n
3
n
2
n
1
n
0
Hex n
7
n
6
n
5
n
4
0 0 0 0 0
1 0 0 0 1
2 0 0 1 0
3 0 0 1 1
4 0 1 0 0
5 0 1 0 1
6 0 1 1 0
7 0 1 1 1
8 1 0 0 0
9 1 0 0 1
A 1 0 1 0
B 1 0 1 1
C 1 1 0 0
D 1 1 0 1
E 1 1 1 0
F 1 1 1 1
n
7
n
6
n
5
n
4
Hex n
11
n
10
n
9
n
8
0 0 0 0 0
1 0 0 0 1
2 0 0 1 0
3 0 0 1 1
4 0 1 0 0
5 0 1 0 1
6 0 1 1 0
7 0 1 1 1
8 1 0 0 0
9 1 0 0 1
A 1 0 1 0
B 1 0 1 1
C 1 1 0 0
D 1 1 0 1
E 1 1 1 0
F 1 1 1 1
n
11
n
10
n
9
n
8
Hex n
15
n
14
n
13
n
12
0 0 0 0 0
1 0 0 0 1
2 0 0 1 0
3 0 0 1 1
4 0 1 0 0
5 0 1 0 1
6 0 1 1 0
7 0 1 1 1
8 1 0 0 0
9 1 0 0 1
A 1 0 1 0
B 1 0 1 1
C 1 1 0 0
D 1 1 0 1
E 1 1 1 0
F 1 1 1 1
n
15
n
14
n
13
n
12
$GFDTA
2
122.8
99
2350
8835
2010/06/10
10:21:25
HF-1015
2800
*
2F
Header
Channel
Number
ppmm
R
2
Distance
m x 10
Light
Level
Date
Time
Serial
Number
Status
Code
End of
String
Check
Sum
$GFDTA, 2, 122.8, 99, 2350, 8835, 2010/06/10 10:21:25, HF-1015,2800 * 2F
Summary of Contents for GasFinderMC-2
Page 8: ...viii Boreal Laser Inc GasFinderMC Operation Manual Contents July 17th 2017 ...
Page 24: ...3 14 Boreal Laser Inc GasFinderMC Operation Manual Installation July 17th 2017 ...
Page 30: ...4 6 Boreal Laser Inc GasFinderMC Operation Manual Data I O July 17th 2017 ...
Page 71: ...BOREAL LASER INC MC Instrument Overall Dimensions JB0125MC A 1 of 1 NTS A Dim mm ...
Page 73: ......
Page 75: ......
Page 79: ......
Page 81: ......
Page 83: ......
Page 85: ......
Page 87: ......
Page 89: ......